articles This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/bync/3.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Copyright © KIChE. All rights reserved.

All issues

Indirect Electrochemical Sensing of the L-abrine Biomarker of Abrin Toxin Via Enzymatic Oxidation of N-methyl-L-tryptophan

Department of Chemical Engineering, Kyonggi University
jjh1023@kyonggi.ac.kr
Korean Journal of Chemical Engineering, August 2026, 43(10), 2625-2633(9)
https://doi.org/10.1007/s11814-026-00732-0

Abstract

The biotoxin abrin, if inhaled or ingested, can be fatal to humans within several days. Because of the initial flu-like 

symptoms and the lack of analytical technologies for determining abrin poisoning, the high possibility of misdiagnosis 

makes this biochemical weapon particularly threatening. This urged us to develop a sensing device for real time monitoring

of abrin poisoning. We introduce an enzyme-linked electrochemical sensor for the detection of the chemical marker 

of abrin poisoning, known as N-methyl-L-tryptophan (MT) or more commonly referred to as L-abrine. Indium tin oxide 

(ITO) was employed as the electrode substrate for immobilization of the enzyme methyl tryptophan oxidase (MTOX), 

which is selective for the MT substrate. The MTOX-immobilized ITO electrode obtained via the amino-glutaraldehyde 

cross-linking chemistry exhibited a sensitivity of 23.3 ± 1.28 µA per decade concentration of MT in a linear detection 

range from 98.1 pM to 1.17 µM (pH 7 in phosphate-buffered saline), and the detection limit was 22.7 pM. Each of the 

electrode modification steps were verified by electrochemical impedance spectroscopy, cyclic voltammetry, and Fourier 

transform-infrared spectroscopy. The approximate amount of MTOX immobilized on a 1 cm2

 ITO substrate was determined

by cyclic voltammetry in a phosphate buffered saline solution at neutral conditions, revealing 3.3 ± 0.46 µC which 

is equivalent to 11.4 pmole of MTOX. With this MT detection strategy in hand, we look to further optimize our approach 

toward a rapid clinical diagnostic tool.