Issue
Korean Journal of Chemical Engineering,
Vol.26, No.2, 496-499, 2009
Effect of laser beam focusing point on AFM measurements
The optical beam deflection method, which is used in AFM to obtain surface images, may distort the resulting image. The flexible and long cantilever is easily overdamped by the laser radiation pressure, resulting in steady deflection of the cantilever (<1 nm). This deflective force distorts the image and influences the force-distance (F-D) curve. The present study investigated the effect of laser radiation pressure on image distortion. As a proof-of-concept test, two grating samples (with step heights of 150 and 18 nm for TGX01 and TGZ01, respectively) were examined with an NSC36 series cantilever in air and water media.
[References]
  1. Kim Y, Kang SK, Choi I, Lee J, Yi J, Appl. Phys. Lett., 88, 173121, 2006
  2. Choi I, Yi J, Korean J. Chem. Eng., 25(2), 386, 2008
  3. Kim Y, Choi I, Kang SK, Lee J, Yi J, Appl. Phys. Lett., 86, 073113, 2005
  4. Kim Y, Yi J, J. Phys. Chem. B, 110(41), 20526, 2006
  5. Hutter JL, Bechhoefer J, Rev. Sci. Instrum., 64, 1868, 1993
  6. Jones RE, Hart DP, Tribol. Int., 38, 335, 2005
  7. Sun Y, Pang JHL, Nanotechnology, 17, 933, 2006
  8. Pal S, Ghosh AK, Electronic Lett., 42, 580, 2006
  9. Dragoman D, Dragoman M, Appl. Opt., 38, 6773, 1996
  10. Kwon J, Honh J, Kim YS, Lee DY, Lee S, Park S, Rev. Sci. Instrum., 74, 4378, 2003
  11. Argento C, French RH, J. Appl. Phys., 80, 6081, 1996
  12. Ackler HD, French RH, Chiang YM, J. Colloid Interface Sci., 179(2), 460, 1996
  13. Carpick RW, Ogletree DF, Salmeron M, J. Colloid Interface Sci., 211(2), 395, 1999